22
Wait
Important note: the unit must not be used during this period.
After twelve seconds the display will change to:
0-6ppm
Wait
The subject may now seal their lips around the mouthpiece and exhale
slowly and fully if they cannot hold their breath any longer. However,
more accurate results will be obtained if they hold their breath until
the buzzer sounds and the display changes to:
Blow
0-6ppm
The CO reading will rise to a plateau over the course of several
seconds. The final value will be held until the unit is turned off and
will be displayed in one of the four ranges:
0–6 ppm with green indicator
7–10 ppm with amber indicator
11–20 ppm with red indicator
20+ ppm with flashing red indicator and alarm